The Mastersizer 2000 has been developed to meet industry's growing need for global comparability of results, traceability, regulatory compliance, and efficiency in the laboratory. In introducing the Mastersizer 2000, Malvern has advanced particle size analysis to the point where it is now a simple, straightforward and routine task.
Accuracy and Reproducibility
Accuracy: ± 1% on the Dv50 using the Malvern Quality Audit Standard. Instrument-to-instrument reproducibility: Better than 1% RSD on the Dv50 using the Malvern Quality Audit Standard.
Assured reproducibility
Software-driven SOPs eliminate user variability and enable global method transfer. All measurement parameters are automatically embedded in the result files and can be critically reviewed by e-mail recipients. Measurements can be replicated by following the same SOP.
Broad measuring range
Measures materials from 0.02µm to 2000µm.
Wide range of sample types
For the measurement of emulsions, suspensions and dry powders.
Ease of use
Fully automated for ease of use. Designed to eliminate user-to-user variability. Reduces training requirements for new users and maximizes deployment of skilled personnel.
Flexibility
Wide range of sample dispersion units. Rapid changeover between sample dispersion units with automatic configuration. "Plug and Play" cassette system allows simultaneous connection of two sample dispersion units.
Regulatory Compliance
Full QSpec validation documentation available including compliance with 21 CFR Part 11.
User-friendly Software
Software-driven Standard Operating Procedures (SOPs) eliminate user variability.
An SOP-creation wizard helps users to create good measurement methods.
On-line help and advice on every aspect of performing a measurement available at the click of a button.
On screen instructions lead user through the measurement process.
SOPs supplied for routine sample analyzes.
Custom Report Designer allows screen layout and printouts to be configured to suit each user.
Custom calculator allows the parameters important to your application to be defined.
Mastersizer 2000 datasheet
The Mastersizer 2000 is a modular instrument designed for the measurement of the particle size distribution of wet and dry samples. It has a wide range of sample dispersion units including an Autosampler. Fully automated, it delivers results based on standardized procedures designed to eliminate user-to-user variability.
Specification
Size range:
0.02 to 2000 microns, depending on material properties.
Measurement principle:
Mie scattering.
Detection systems:
Red light: Forward scattering, side scattering and back scattering.
Blue light: Wide angle forward and back scattering.
Light sources:
Red light: Helium neon laser.
Blue light: Solid state light source.
Optical alignment system:
Automatic rapid align system with dark field optical reticle and multi-element alignment detector.
Sample dispersion unit interchange system:
Sample dispersion units automatically recognized, configured and enabled on insertion of measurement cell cassettes into optical unit.
Power:
110/240V, 50/60 Hz, 60 VA.
Dimensions:
Length: 1293 mm.
Depth: 255 mm.
Height: 375 mm.
Weight:
31 kg.
Chemical compatibility
It should be noted that no sample or dispersant should come into contact with the Mastersizer 2000 optical unit in normal operation as all sample and dispersant is contained within the dispersion accessories and the sample cell.