Laser diffraction.
|
|
|
| |
0.1µm to 2000µm
|
|
|
| |
300 mm lens: 0.1 – 900 microns (Dv50 : 0.5 – 600 microns).
750 mm lens: 2 – 2000 microns (Dv50 : 5 – 1600 microns).
|
|
|
| |
0.5 microns: 150mm
5 microns: > 1000mm
10 microns: > 2000mm
|
|
|
| |
Mie Theory and Fraunhofer Approximation .
|
|
|
| |
Patented Multiple Scattering Correction ( US Patent No. 5,619,324) enables operation at high concentrations.
Minimum acceptable transmission: 5% (95% obscuration), dependent on particle size range.
|
|
|
| |
36 element log-spaced silicon diode detector array
|
|
|
| |
632.8 nm, 2 mW helium-neon laser.
|
|
|
| |
Class 3R laser product.
|
|
|
| |
Automatic rapid align system
|
|
|
| |
Continuous Mode: 1Hz
Rapid Mode (Standard): 2.5kHz
Rapid Mode (Enhanced): 10kHz (requires software feature key) .
|
|
|
| |
Continuous Mode: > 60 mins at 1Hz
Rapid Mode (Standard): maximum of 120 sec at 2.5 kHz
Rapid Mode (Enhanced): maximum of 30 sec at 10 kHz .
|
|
|
| |
Better than +/- 1% on the Dv50 for NIST-traceable latex standards
|
| |
Better than +/- 1% COV on the Dv50 for NIST-traceable latex standards
|
| |
Better than +/- 1% COV on the Dv50 for NIST-traceable latex standards |
|
|
|
|
| |
Internal: Based on transmission or light scattering levels
External: Based on TTL input or simple switch trigger
|
|
|
| |
Via two TTL trigger outputs
|
|
|
|
|
| |
Pentium PC 2GHz, 128MByte RAM, 80MByte free hard disk space, 1024 x 768 screen resolution, CD-ROM drive, Windows 2000 ( Service pack 3 or higher) or XP professional (Service pack 1a or higher). One free USB port.
|
|
|
| |
Windows XP Professional (service pack 1 or higher). |
|
|
| |
Automated using Standard Operating Procedures created in the software.
Manual using on-screen controls and hot keys.
|
|
|
| |
Model independent analysis to provide accurate droplet size distributions.
Rosin-Rammler and Log-Normal 2-parameter models available.
|
|
|
| |
User-configurable Particle Size History display, enabling frame-by-frame data analysis.
|
|
|
| |
Searching, sorting and filtering by search criteria of data records on all parameters of interest.
|
|
|
| |
Direct to Excel or other spreadsheet packages.
|
|
|
| |
Technical compliance provided through layered access security system, audit trails and links to Adobe Acrobat for electronic signatures.
|
|
|
|
|
| |
IP65 rating for the main transmitter and receiver modules in normal use.
|
|
|
| |
Air purge system available to prevent optics contamination when measuring wide spray plumes. Requires a continuous supply of clean, compressed air at a flow rate of at least 80 l/min.
|
|
|
| |
15ºC – 40°C.
|
|
|
| |
35% - 80%, non-condensing.
|
|
|