Summary
The Parsum probe is inserted directly into a process line with the following general configuration and parameters.
|
|
|
General |
|
|
Intrinsic safety |
The IPP 50 unit is not Intrinsically Safe (See IPP 50-Se)
|
|
|
Sensor |
Applies an extended fibre-optic local filter method
|
|
|
Installation
|
Minimal installation required. Is inserted directly into process line. Usually retrofitted easily in pipes, vessels and reactors. |
|
|
Process interface kit
|
- Weld Stub to be installed on the process line
- Feed-thru piece to hold probe
- Tri-clamp to hold probe to the feed-thru piece
|
|
|
Air supply
|
No air is required for the IPP 50 unit . |
|
|
Electronics Enclosure
|
There is no electronics enclosure for the IPP 50 unit. |
|
|
Cables
|
- 50m of cable is supplied with the system.
|
|
|
Site Requirements
|
- Installation location to be specified following consultation of a Malvern Process Specialist
- Welding the support stub in the process line
- Conduit laid for cables to computer and to control room
- Minimum 120 V AC 5amp power supply
|
|
|
| Software and Hardware |
- Inline Particle Probe Software V. 5.54
- NI 5102 A/D Converter card
- Field mounted computer with Windows 2000 operating system
|
|
|
| Feedback Control Options |
- PCI -6308A D/A Converter Card (4-20 mA Outputs)
- RS485 Interface Card
|
|
|
Miscellaneous |
|
|
Installation & Training |
|
|
|
Standard documentation |
The following documents are supplied on a CD-ROM in PDF format and are available in English, German, French and Spanish languages
- Commissioning Manual
- Operator Control Software Manual
|
|
|
Specification |
|
|
| Measurement range |
- Particle Size Measuring Range 50-2500µm.
- Velocity Measurement Range 0.01-20m/s.
|
|
|
| Intrinsically safe |
The IPP 50 unit is not intrinsically safe.
|
|
|
| Materials |
- 316L SS for in-line probe.
- Sapphire, epoxy resin optics.
- Pressure-cast aluminum for casing.
|
|
|
| Type of measurement material |
Solid, liquid and gaseous particles.
|
|
|
| Concentration |
200-120,000 ppm volume content, dependent on particle size and velocity of the moving particles.
|
|
|
| Data rate |
Up to 500 measurements/second (depending upon the process).
|
|
|
| Air Requirements |
The IPP 50 unit does not require compressed air.
|
|
|
| Maximum Operating Pressure |
4 bar at the probe's sensing volume.
|
|
|
| Operating Temperatures |
- -20°C to 85°C at the measuring point
- -10°C to 60°C on the probe housing
|
|
|
| Industrial Protection |
IP 65 (probe)
|
|
|
| Maximum Number of Instruments Connected to one PC |
4
|
|
|
| Dimensions |
- Tube Length = 280mm (~11in).
- Tube Diameter = 25mm (~ 1in).
|
|
|
| Weight (approx) |
1.5kg
|
|
|
| |